P25 Phase 2 Commands
:SOURce:FREQuency:CENTer <value>
:SOURce:FREQuency:CENTer?
Description: Sets the signal generator center frequency. Please note that setting the
Cmd Parameters: <value>
Query Parameters: NA
Range: 500000 Hz to 1600000000 Hz
Default Value: 800000000 Hz
Default Unit: Hz
Example: To set the signal generator center frequency to 145 MHz:
Front Panel Access: Frequency, Tx Freq
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <value>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Description: Sets the output power level for the P25p2 signal generator. Please note
Cmd Parameters: <value>
Query Parameters: NA
Range: 0 dBm to -130 dBm or 1 mW to 1 fW or
Default Value: 0 dBm or 1 mW or 223606797749978 fV
Default Unit: dBm or fW or fV
Example: To set the signal generator output level to -10 dBm:
Front Panel Access: Amplitude, Tx Output Lvl
S412E PM
center frequency will restart the sweep. The query returns the current
signal generator frequency in Hz.
:SOURce:FREQuency:CENTer 145000000
that changing the Tx power offset will also cause the display of this to
adjust to the new offset. For example, if the output level is set to 0 dBm
and the Tx level offset is then set to 10 dB external gain, the max limit
and value of the Tx output level will be adjusted to 10 dBm. The query
returns the current Tx output level.
The query will be returned in the unit that is selected through the
Tx Units button on the front panel or with the command:
UNIT:POWer:TX. The set command must be sent using the units
selected. If the receiver unit has been set to dBm, the generator output
level is returned in dBm and must be set in dBm. If the unit is set to
Watts, the generator output level is returned in fW (10
be set in fW. If the unit is set to Volts, the generator output level is
-15