5-2 :CALCulate Subsystem
:CALCulate<Tr>:LIMit:LOWer:Y <y-parameter>
:CALCulate<Tr>:LIMit:LOWer:Y?
Description: Sets the location of the lower limit line of the given trace <Tr> on the
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Units: Current active trace y-axis unit
Related Command: :CALCulate:LIMit[:STATe]
Front Panel Access: Shift 6 (Limit), Limit Edit, Amplitude
:CALCulate:LIMit:POINt?
Description: Returns the number of points currently in the selected limit line.
Cmd Parameter: NA (query only)
Query Response: <NR1> <integer>
Related Command: :CALCulate:LIMit:TYPE
Front Panel Access: NA
:CALCulate:LIMit:POINt:ADD
Description: Adds a new limit point to the currently active limit line. Use
Cmd Parameter: NA
Query Response: NA (no query)
Related Command: :CALCulate:LIMit:TYPE
Front Panel Access: Shift 6 (Limit), Limit Edit, Add Point
5-10
y-axis at the given value. This moves the entire lower limit and moves
the current active limit point by the given value. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. The <y-parameter> is
defined in the current y-axis. If no unit is specified with the
<y-parameter>, then the default unit is used. The set version of the
command changes the Move Limit on the front panel to Limit if it is
currently set to Point, and sets the given trace as the active trace. The
query version of the command returns the location of the active limit
point on the y-axis. If an error occurs, such as limit not ON, then the
query version of the command returns a –400 error codes. Limit line
must be ON for the command to be valid.
:CALCulate3:LIMit:LOWer:X
Use :CALCulate:LIMit:TYPE to set the currently active limit line.
:CALCulate:LIMit:TYPE to set the currently active limit line.
:CALCulate:LIMit:POINt:DELete
PN: 10580-00319 Rev. N
VNA Commands
S412E PM