VNA Commands
Related Command: :CALCulate:LIMit:TYPe
Front Panel Access: Shift 6 (Limit), Limit Edit, Limit X
:CALCulate:LIMit:Y <y-parameter>
:CALCulate:LIMit:Y?
Description: Sets the location of the active limit line on the y-axis at the given value.
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Units: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPe
Front Panel Access: Shift 6 (Limit), Limit Edit, Amplitude
S412E PM
OR to 500 MHz:
:CALCulate:LIMit:X 500MHz
To move the active limit to 5 Feet (active trace in distance domain with
current distance unit in feet):
:CALCulate:LIMit:X 5FT
OR to 4 Meter
:CALCulate:LIMit:X 4M
:CALCulate:LIMit:X 4
:CALCulate:LIMit:Y
This moves the entire limit and moves the current active limit point by
the given value. Sending this command will change the Move Limit on
the front panel to Limit if it is currently set to Point. The
<y-parameter> is defined in the current y-axis. If no unit is specified
with the <y-parameter> then the default unit will be used. The query
version of the command returns the location of the active limit point on
the y-axis. If an error occurs, such as limit not ON, the query version of
the command returns an error code of –400. Limit line must be on for
the command to be valid. Use :CALCulate:LIMit:TYPE to set the
currently active limit line.