5-2 :CALCulate Subsystem
:CALCulate:LIMit:TYPE 0|1
:CALCulate:LIMit:TYPE?
Description: Sets the limit line segment type (upper or lower) to be edited. Set the
Cmd Parameter: <char> 0|1
Query Response: <char> 0|1
Default Value: 0
Example: To set upper limit line active for editing:
Front Panel Access: Shift 6 (Limit), Limit
:CALCulate<Tr>:LIMit:UPPer:POINt?
Description: Query only. Returns the number of points currently in the upper limit
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt?
Cmd Parameter: NA (query only)
Query Response: <NR1> <integer>
Example: To query for the upper limit line total point on trace #2:
Front Panel Access: NA
:CALCulate<Tr>:LIMit:UPPer:POINt:ADD
Description: Adds a new limit point to the upper limit line of the given trace <Tr>.
Parameter: NA
Query Syntax: NA (no query)
Query Response: NA
Example: To add a point to the upper limit line on trace 2:
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:DELete
Front Panel Access: Shift 6 (Limit), Limit Edit, Add Point
5-14
value to 1 for Lower limit segment and to 0 for Upper limit line
segment. The query version of the command returns a 1 if the lower
limit line is currently active for editing and returns a 0 if the upper
limit line is currently active for editing.
:CALCulate:LIMit:TYPE 0
To set lower limit line active for editing:
:CALCulate:LIMit:TYPE 1
line of the given trace <Tr>. <Tr> is the trace number in the range
1 to 4. If no trace number is specified, then the <Tr> parameter
defaults to trace number 1.
:CALC2:LIM:UPP:POIN?
<Tr> is the trace number in the range 1 to 4. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1.
:CALC2:LIM:UPP:POIN:ADD