Anritsu LMR Master S412E Programming Manual page 106

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

4-14 [:SENSe] Subsystem
[:SENSe]:OBWidth:XDB <rel ampl>
[:SENSe]:OBWidth:XDB?
Title: Occupied Bandwidth dB Down
Description: This command sets the number of dB below the carrier used to measure
Parameter: <rel ampl>
Default Value: 3 dBc
Default Unit: dBc
Range: 0 to 100 dBc
Related Command: :OBWidth:METHod
Front Panel Access: Shift-4 (Measure), OCC BW, dBc
[:SENSe]:OPTion:IF:BWSTate 1|2|3|4|5
:SENSe:OPTion:IF:BWSTate?
Title: IF Bandwidth State
Description: Sets the IF Bandwidth. Setting a value of 1 is equivalent of setting the
Parameter: 1|2|3|4|5
Range: 1|2|3|4|5
[:SENSe]:POWer[:RF]:ATTenuation <rel ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Title: Input Attenuation
Description: Sets the input attenuation. Note that issuing this command will set the
Parameter: <rel ampl>
Default Value: 30 dB
Default Unit: dB
Range: 0 dB to 65 dB
Related Command: :POWer[:RF]:ATTenuation:AUTO
Front Panel Access: Amplitude, Atten Lvl
4-64
the occupied bandwidth. This value is used in the measurement if
:SENSe:OBWidth:METHod is set to XDB.
IF Bandwidth to Normal on the front panel. Setting a value of 3 is
equivalent of setting the IF Bandwidth to 10 MHz BW on the front
panel. Setting the value of 4 is equivalent of setting the IF Bandwidth to
16 MHz BW on the front panel. Note that option 89 must be available
and in Zero span for the command to be valid. The query version of this
command will return a 0 if not in zero span.
automatic input attenuation OFF.
PN: 10580-00319 Rev. N
Spectrum Analyzer Commands
S412E PM