Read Subsystem - Anritsu LMR Master S412E Programming Manual

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

NXDN Commands

12-9 :READ Subsystem

This set of commands combines the ABORt, INITiate and FETCh commands. It aborts any
current triggering sequence and sets the trigger state to idle. It then initiates a new active
measurement (i.e., begins the collection of new data). When the measurement is complete,
it returns the result. These commands will not switch to another measurement.
To prepare for a new measurement, use the CONFigure command. To get the current
measurement data, use the FETCh command.
:READ:COVerage?
Description: Triggers a new NXDN Coverage measurement and returns the
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To read NXDN Coverage numerical data:
Related Command: :CONFigure:COVerage
Front Panel Access: NA
S412E PM
numerical results. It is a combination of the commands :ABORT;
:INITiate; :FETCh:COVerage? NXDN Coverage must be the active
measurement (specified by :CONFigure:COVerage). The current
measurement can be queried using :CONFigure?
Data is returned as 7 comma-separated values in the following order
and format:
RSSI (dBm as float)
BER (% as float)
Mod Fid (% as float)
Latitude (radian as float)
Longitude (radian as float)
UTC Date (month/day/year as char)
UTC Time (hour:minute:second as char)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--".
:READ:COVerage?
PN: 10580-00319 Rev. N
12-9 :READ Subsystem
12-19

Advertisement

Table of Contents
loading

Table of Contents