Anritsu LMR Master S412E Programming Manual page 124

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

5-2 :CALCulate Subsystem
:CALCulate<Tr>:LIMit:UPPer:POINt:RIGHt
Description: Sets the limit point to the right of the upper limit active point of the
Parameter: NA
Query Syntax: NA (no query)
Query Response: NA
Example: To make the upper limit point to the right of the current active point of
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:LEFT
Front Panel Access: Shift 6 (Limit), Limit Edit, Next Point Right
:CALCulate<Tr>:LIMit:UPPer:POINt:X <x-parameter>
:CALCulate<Tr>:LIMit:UPPer:POINt:X?
Description: Sets the location of the upper limit point of the given trace <Tr> on the
Cmd Parameter: <NRf> <x-parameter> (hertz, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, meters, feet)
Default Units: Hz for Frequency domain,
Example: To set the trace 4 upper limit point to 5000 Hertz (trace 4 in frequency
5-16
given trace <Tr> as the new active point. <Tr> is the trace number in
the range 1 to 4. If no trace number is specified than defaults to trace
number 1.
trace 2 as the new active point:
:CALCulate2:LIMit:UPPer:POINt:RIGHt
:CALC2:LIM:UPP:POIN:RIGH
x-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified than defaults to trace number 1.
Sending the set command will change the Move Limit on the front panel
to Point if it is currently set to Limit and set the given trace as the
active trace. <x-parameter> is defined in the given trace current x-axis.
The given unit must correspond to the given trace domain type. If no
unit is specified with the <x-parameter> then the default unit will be
used. The query version of the command returns the location of the
given trace upper limit point on the x-axis followed by the unit. If an
error occurs, such as limit not ON, the query version of the command
returns an error code of –400. Limit line must be on for the command to
be valid. Use :CALCulate:LIMit:TYPe to set the currently active
limit line.
Meters or Feet for distance domain.
domain):
:CALCulate4:LIMit:UPPer:POINt:X 5000
OR to 500 MHz:
:CALCulate4:LIMit:UPPer:POINt:X 500 MHz
PN: 10580-00319 Rev. N
VNA Commands
S412E PM