Spectrum Analyzer Commands
[:SENSe]:DETector[:FUNCtion] POSitive|RMS|NEGative|SAMPle
[:SENSe]:DETector[:FUNCtion]?
Title: Detection Type
Description: Sets the detection method for calculating each display point. Each
Parameter: POSitive|RMS|NEGative|SAMPle
Parameter Type: <char>
Default Value: (Positive) Peak
Front Panel Access: Amplitude, Detection
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Title: Center Frequency
Description: Sets the center frequency. Note that changing the value of the center
Parameter: <freq>
Default Unit: Hz
Front Panel Access: Freq, Center Freq
[:SENSe]:FREQuency:SIGStandard:CHANnel <number>
[:SENSe]:FREQuency:SIGStandard:CHANnel?
Title: Channel Selection
Description: Sets the channel number for the selected signal standard.
Parameter: <number>
Front Panel Access: Freq, Channel
S412E PM
display point represents several measurements. The detection type
determines how the display point is derived from its associated
measurements. POSitive Peak detection displays the maximum value of
the associated measurements. RMS detection displays the average
power of the associated measurements. NEGative Peak detection
displays the minimum value of the associated measurements. SAMPle
detection displays the "middle" point of those measurements associated
with a display point. For example, if there are 3 measurement
frequencies associated with a given display point, sample detection will
display the value at the frequency of the second measurement point.
frequency will change the value of the coupled parameters Start
Frequency and Stop Frequency. It may also change the value of the
span.