11-9 :READ Subsystem
:READ:SIGAnalyzer?
Description: Triggers a new P25p2 Analyzer measurement and returns the
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To read P25p2 Analyzer numerical data:
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: NA
11-22
numerical results. It is a combination of the commands :ABORT;
:INITiate; :FETCh:SIGAnalyzer?
P25p2 Analyzer must be the active measurement (specified by
:CONFigure:SIGAnalyzer). The current measurement can be queried
using :CONFigure? Data is returned as 7 comma-separated values in
the following order and format:
Received Power (dBm as float or Watts as long long int or Volts as
long long int)
Frequency Error (Hz as float)
Mod Fid (% as float)
BER (% as float)
Symbol Dev (Hz as float)
CC (hex)
Symbol Rate Error (Hz as float)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--".
The received power will be returned in the unit that is selected through
the Rx Units button on the front panel or with the command:
UNIT:POWer:RX. If the receiver unit has been set to dBm, the received
power is returned in dBm. If the unit is set to Watts, the received power
is returned in fW (10
is returned in fV (10
Note: This command is not affected by the squelch level set using the
front panel.
:READ:SIGAnalyzer?
PN: 10580-00319 Rev. N
-15
W). If the unit is set to Volts, the received power
-15
V).
P25 Phase 2 Commands