7-2 :CONFigure Subsystem

7-2
:CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement. It sets the
instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the
taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
:CONFigure SUMMary
Title: Configure Summary Measurement
Description: This command configures the summary of all the related numerical
Parameter: SUMMary
Parameter Type: <char>
Front Panel Access: Shift-4 (Measure), WiMAX Summary
:CONFigure:DEMod <char>
Title: Configure Demodulation Measurement
Required Option: 47
Description: Valid <char> parameters are:
7-2
measurement results. It disables any other active measurements. None
of the instrument parameters are changed by the execution of this
command. Sets the instrument to single sweep mode
(:INITiate:CONTinuous OFF). Measurement settings can be modified
by using the [:SENSe] commands before initiating a measurement.
SUMMary|CONStln|SFLatness|EVSCarrier|EVSYmbol.
This command configures the selected demodulation measurement. It
disables any other active measurements. None of the instrument
parameters are changed by the execution of this command. Sets the
instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]:DEMod
commands before initiating a measurement.
PN: 10580-00319 Rev. N
Fixed WiMAX Commands
S412E PM