NBFM Commands
:FETCh:SIGAnalyzer?
Description: Returns the most recent NBFM Analyzer numerical measurement
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To fetch NBFM Analyzer numerical data:
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: NA
S412E PM
results. Data is returned as 10 comma-separated values in the following
order and format:
Carrier Power (dBm as float or Watts as long long int or Volts as
long long int)
Carrier Frequency (Hz as float)
Deviation (Hz as float)
Mod Rate (Hz as float)
SINAD (dB as float)
THD (% as float)
Occ BW (Hz as float)
ExtSINAD (dB as float)
CTCSS (string)
DCS (string)
DTMF (string)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--,--,--,--,--".
Please note that the received power will be returned in the unit that is
selected through the Rx Units button on the front panel or with the
command :UNIT:POWer:RX. If the receiver unit has been set to dBm,
the received power is returned in dBm. If the unit is set to Watts, the
received power is returned in fW (10
received power is returned in fV (10
Also note that only one of the last three numerical values will be shown.
The other two values will be dashed out. This is done because only one
tone type can be chosen at any given time.
:FETCh:SIGAnalyzer?
PN: 10580-00319 Rev. N
16-4 :FETCh Subsystem
-15
W). If the unit is set to Volts, the
-15
V).