5-2 :CALCulate Subsystem
Front Panel Access: Marker, Marker Type
:CALCulate:MARKer<Mk>:SOURce <Tr>
:CALCulate:MARKer<Mk>:SOURce?
Description: Sets the specified marker <Mk> to the given trace <Tr>. <Mk> is the
Cmd Parameter: <char> <Tr>
Query Response: <char> <Tr>
Front Panel Access: Marker, Marker on Trace
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:TYPE
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:TYPE?
Description: Sets the specified marker to the given marker type and make it the
Cmd Parameter: <char> REFerence|DELTa|OFF
Query Response: <char> REF|DELT|OFF
Default Value: OFF
Example: To set marker #1 as the reference marker and turn it on:
Front Panel Access: Marker, Marker Type
5-26
To turn off marker #6:
:CALCulate:MARKer6:REFerence OFF
:CALCulate:MARKer6:REFerence:STATe OFF
:CALCulate:MARKer6:REFerence:STATe 0
marker number in the range of 1 to 8. If no marker number is specified,
then the marker number (the <Mk> value) defaults to 1. <Tr> is the
trace and must be one of the following 9 values:
TR1|TR2|TR3|TR4|MEM1|MEM2|MEM3|MEM4|ALL
The query version of the command returns "TR1" if the specified marker
is on trace 1, "TR2" if on trace 2, "TR3" if on trace 3, "TR4" if on trace 4,
"MEM1" if on trace 1 memory, "MEM2" if on trace 2 memory, "MEM3" if
on trace 3 memory, "MEM4" if on trace 4 memory, and "ALL" if the
specified marker is on all 4 traces. Note that the set version of this
command will set the specified marker as the active marker.
If an error occurs, such as "Marker not ON", the query version of the
command returns an error code of –400.
active marker. If set to REF then the specified marker is turn on and is
set as reference marker. If set to DELT then the specified marker is
turn on and is set as a delta marker. If set to OFF then the specified
marker is turn off. The query version of this command returns the
string "REF" if the specified marker is set as reference marker, "DELT"
if set as delta marker, or "OFF" if the specified marker is currently set
to off.
:CALCulate:MARKer1:TYPE REFerence
:CALCulate:MARKer:TYPE REF