Anritsu LMR Master S412E Programming Manual page 120

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

5-2 :CALCulate Subsystem
:CALCulate:LIMit:POINt:X <x-parameter>
:CALCulate:LIMit:POINt:X?
Description: Sets the location of the active limit point on the x-axis at the specified
Cmd Parameter: <NRf> <x-parameter> (hertz, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, meters, feet)
Default Units: Hz for Frequency domain,
Example: To set the active limit point to 5000 Hertz (active trace in frequency
Related Command: :CALCulate:LIMit:POINt:Y
Front Panel Access: Shift 6 (Limit), Limit Edit, Limit X
5-12
location. Sending this command changes the Move Limit on the front
panel to Point if it is currently set to Limit. The <x-parameter> must
correspond to the current active trace domain type. If no unit is
specified with the <x-parameter>, then the default unit is used. The
query version of the command returns the location of the active limit
point on the x-axis followed by the unit. If an error occurs, such as limit
not ON, then the query version of the command returns –400 error
codes. Limit line must be ON for the command to be valid. Use
:CALCulate:LIMit:TYPE to set the currently active limit line.
Meters or Feet for distance domain.
domain):
:CALCulate:LIMit:POINt:X 5000
OR to 500 MHz:
:CALCulate:LIMit:POINt:X 500 MHz
To set the active limit point to 5 Feet (active trace in distance domain
with current distance unit in meter):
:CALCulate:LIMit:POINt:X 5 FT
OR to 4 Meter
:CALCulate:LIMit:POINt:X 4 M
:CALCulate:LIMit:POINt:X 4
PN: 10580-00319 Rev. N
VNA Commands
S412E PM