Spectrum Analyzer Commands
:TRACe1:OPERation NORMal|MAXHold|MINHold|AVERage
:TRACe1:OPERation?
Title: Trace A Operation
Description: This command specifies how successive traces are combined to produce
Parameter: NORMal|MAXHold|MINHold|AVERage
Parameter Type: <char>
Default Value: NORMal
Related Command: :AVERage:TYPE
:TRACe2:OPERation MAXHold|MINHold
:TRACe2:OPERation?
Title: Trace B Operation
Description: This command specifies how successive traces are combined to produce
Parameter: MAXHold|MINHold
Parameter Type: <char>
Default Value: None
Range: MAXHold|MINHold
S412E PM
the resulting display values.
Setting the operation to NORMal is equivalent to pressing the Shift-5
(Trace), Trace A Operations, Normal->A on the front panel. This
displays a trace based on the detection method selected.
Setting the operation to MAXHold is equivalent to pressing the Shift-5
(Trace), Trace A Operations, Max Hold->A on the front panel. This
displays the largest signal for each display point over multiple sweeps.
Setting the operation to MINHold is equivalent to pressing the Shift-5
(Trace), Trace A Operations, Min Hold->A on the front panel. This
displays the smallest signal for each display point over multiple sweeps.
Setting the operation to AVERage is equivalent to pressing the Shift-5
(Trace), Trace A Operations, Average->A on the front panel. This
displays the average value of multiple sweeps for each display point.
The query version of the command returns the current operation mode
or "NONE" if no operation is set.
the resulting display values.
Setting the operation to MAXHold is equivalent to pressing the Shift-5
(Trace), Trace B Operations, Max Hold->B on the front panel. This
displays the largest signal for each display point over multiple sweeps.
Setting the operation to MINHold is equivalent to pressing the Shift-5
(Trace), Trace B Operations, Min Hold->B on the front panel. This
displays the smallest signal for each display point over multiple sweeps.
The query version of the command returns the current operation mode
or "NONE" if no operation is set.