10-13 [:SENSe] Subsystem
[:SENSe]:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Description: Turns auto range for the receiver on or off. When auto range is on, the
Cmd Parameters: OFF|ON|0|1
Query Parameters: NA
Range: OFF|ON|0|1
Default Value: ON or 1
Default Unit: NA
Example: To turn auto ranging off:
Front Panel Access: Amplitude, Auto Rx Range
[:SENSe]:SYMBolspan <value>
[:SENSe]:SYMBolspan?
Description: Sets the symbol span. Please note that this setting only affects the
Cmd Parameters: <value>
Query Parameters: NA
Range: 2 to 5
Default Value: 2
Default Unit: NA
Example: To set the symbol span to 4:
Front Panel Access: Measurement, P25 Analyzer, Symbol Span
10-42
reference level is automatically adjusted to the proper value to show the
trace on the screen. If the auto ranging is turned off, the reference level
will not adjust according to where the trace is. In P25 Analyzer
measurement, this command adjusts the reference level of the spectrum
graph.
:POWer:RANGe:AUTO OFF
Eye Diagram in the P25 Analyzer measurement. Please note that
setting the symbol span will restart the sweep.
:SYMBolspan 4
PN: 10580-00319 Rev. N
P25 Phase 1 Commands
S412E PM