Anritsu LMR Master S412E Programming Manual page 87

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

Spectrum Analyzer Commands
:TRACe{1|2|3}:LOAD:DISPlay[:STATe] OFF|ON|0|1
:TRACe{1|2|3}:LOAD:DISPlay[:STATe]?
Title: Trace View State
Description: Specifies whether the specified recall trace should be displayable
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Example: To set Trace A to Blank:
Front Panel Access: Shift-5 (Trace), Trace [A/B/C], View Blank
These SCPI commands are applicable only if the user selects the option to recall
Note
the trace to Trace A and not show the live trace.
S412E PM
(visible) or hidden. TRACe1 corresponds to Trace A, TRACe2
corresponds to Trace B, and TRACe3 corresponds to Trace C. Setting
the value to ON or 1 will set the specified trace visible. Setting the value
to OFF or 0 will set the specified trace hidden. Note that this command
is valid only if there is a trace recalled.
:TRACe:LOAD:DISPlay OFF
:TRACe1:LOAD:DISPlay 0
To set Trace B to View:
:TRACe2:LOAD:DISPlay ON
:TRACe2:LOAD:DISPlay:STATe ON
:TRACe2:LOAd:DISPlay 1
PN: 10580-00319 Rev. N
4-11 :TRACe Subsystem
4-45

Advertisement

Table of Contents
loading