5-2 :CALCulate Subsystem
:CALCulate<Tr>:LIMit:UPPer:Y <y-parameter>
:CALCulate<Tr>:LIMit:UPPer:Y?
Description: Sets the location of the upper limit line of the given trace <Tr> on the
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Units: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPe
Front Panel Access: Shift 6 (Limit), Limit Edit, Amplitude
:CALCulate:LIMit:X <x-parameter>
:CALCulate:LIMit:X?
Description: Sets the location of the active limit point on the x-axis at the specified
Cmd Parameter: <NRf> <x-parameter> (hertz, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, meters, feet)
Default Units: Hz for Frequency domain,
Example: To move the active limit to 5000 Hertz (active trace in frequency
5-20
y-axis at the given value. This moves the entire upper limit and moves
the current active limit point by the given value. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified than defaults
to trace number 1. The <y-parameter> is defined in the current y-axis.
If no unit is specified with the <y-parameter> then the default unit will
be used. The set version of the command will change the Move Limit on
the front panel to Limit if it is currently set to Point and set the given
trace as the active trace. The query version of the command returns the
location of the active limit point on the y-axis. If an error occurs, such as
limit not ON, the query version of the command returns an error code
of –400. Limit line must be on for the command to be valid. Use
:CALCulate:LIMit:TYPE to set the currently active limit line.
:CALCulate<Tr>:LIMit:UPPer:X
location. This moves the entire limit and moves the active limit point to
the given value. The <x-parameter> given unit must correspond to the
current active trace domain type. If no unit is specified with the
<x-parameter> then the default unit will be used. Sending the set
command will change the Move Limit on the front panel to Limit if it is
currently set to Point. The query version of the command returns the
location of the active limit point on the x-axis followed by the unit. If an
error occurs, such as limit not ON, the query version of the command
returns an error code of –400. Limit line must be on for the command to
be valid. Use :CALCulate:LIMit:TYPE to set the currently active
limit line.
Meters or Feet for distance domain.
domain):
:CALCulate:LIMit:X 5000
PN: 10580-00319 Rev. N
VNA Commands
S412E PM