Anritsu LMR Master S412E Programming Manual page 397

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

NXDN Commands
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude] <value>
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]?
Description: Sets the output power level for the NXDN signal generator. Please note
Cmd Parameters: <value>
Query Parameters: NA
Range: 0 dBm to -130 dBm or 1 mW to 1 fW or
Default Value: 0 dBm or 1 mW or 223606797749978 fV
Default Unit: dBm or fW or fV
Example: To set the signal generator output level to -10 dBm:
Front Panel Access: Amplitude, Tx Output Lvl
:SOURce:STATe OFF|ON|0|1
:SOURce:STATe?
Description: Turns the signal generator ON or OFF. Please note that the Generator
Cmd Parameters: OFF|ON|0|1
Query Parameters: NA
Range: OFF|ON|0|1
Default Value: OFF
Default Unit: NA
Example: To turn the signal generator on:
Front Panel Access: Turn Sig-Gen ON/OFF
S412E PM
that changing the Tx power offset will also cause the display of this to
adjust to the new offset. For example, if the output level is set to 0 dBm
and the Tx level offset is then set to 10 dB external gain, the max limit
and value of the Tx output level will be adjusted to 10 dBm. The query
returns the current Tx output level.
The query will be returned in the unit that is selected through the
Tx Units button on the front panel or with the command:
UNIT:POWer:TX. The set command must be sent using the units
selected. If the receiver unit has been set to dBm, the generator output
level is returned in dBm and must be set in dBm. If the unit is set to
Watts, the generator output level is returned in fW (10
be set in fW. If the unit is set to Volts, the generator output level is
-15
returned in fV (10
70710678 fV to 223606797749978 fV
:SOURce:POWer -10
ON/OFF button will toggle depending on the state. When the signal
generator is on, the button will show Turn Sig-Gen OFF. When the
signal generator is off, the button will show Turn Sig-Gen ON. The query
returns the current signal generator state. A return value of 1 means
ON and a return value of 0 means OFF.
:SOURce:STATe ON
PN: 10580-00319 Rev. N
V) and must be set in fV.
12-10 :SOURce Subsystem
-15
W) and must
12-23