Anritsu LMR Master S412E Programming Manual page 235

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

Mobile WiMAX Commands
:CONFigure:DEMod <char>
Title: Configure Demodulation Measurement
Required Option: 47
Description: Valid <char> parameters are:
Parameter: <char>
Front Panel Access: Measurements, Demodulator
:CONFigure:RF <char>
Title: Configure RF Measurement
Required Option: 46
Description: Valid parameter options: SUMMary|SPECtrum|PVTime|ACPR
S412E PM
SUMMary|CONStln|SFLatness|EVSCarrier|EVSYmbol.
This command configures the selected demodulation measurement. It
disables any other active measurements. None of the instrument
parameters are changed by the execution of this command. Sets the
instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]:DEMod
commands before initiating a measurement.
When the SUMMARY option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then Demodulator, then Modulation Summary from the front panel.
When the CONStln option is selected, the measurement results are
equivalent to the results that are displayed by choosing Measurements,
then Demodulator, then Constellation from the front panel. When the
SFLatness option is selected, the measurement results are equivalent to
the results that are displayed by choosing Measurements, then
Demodulator, then Spectral Flatness from the front panel. When the
EVSCarrier option is selected, the measurement results are equivalent
to the results that are displayed by choosing Measurements, then
Demodulator then EVM vs Sub Carrier from the front panel. When the
EVSYmbol option is selected, the measurement results are equivalent
to the results that are displayed by choosing Measurements, then
Demodulator then EVM vs Symbol from the front panel.
This command configures the selected RF measurement. It disables any
other active measurements. None of the instrument parameters are
changed by the execution of this command. Sets the instrument to
single sweep mode (:INITiate:CONTinuous OFF). Measurement
settings can be modified by using the [:SENSe]:RF commands before
initiating a measurement.
PN: 10580-00319 Rev. N
8-2 :CONFigure Subsystem