16-13 [:SENSe] Subsystem
[:SENSe]:FILTer:FREQuency 0.3|3|15|NONE
[:SENSe]:FILTer:FREQuency?
Description: Sets the low pass filter frequency.
Cmd Parameters: 0.3|3|15|NONE
Query Parameters: NA
Range: 0.3|3|15|NONE
Default Value: 3
Default Unit: kHz
Example: To set the low pass filter frequency to 15 kHz:
Front Panel Access: Setup, Filter, Low Pass Filter
[:SENSe]:FILTer:HPASs:FREQuency 0.3|3|NONE
[:SENSe]:FILTer:HPASs:FREQuency?
Description: Sets the high pass filter frequency.
Cmd Parameters: 0.3|3|NONE
Query Parameters: NA
Range: 0.3|3|NONE
Default Value: None
Default Unit: kHz
Example: To set the high pass filter frequency to 300 Hz:
Front Panel Access: Setup, Filter, High Pass Filter
16-46
Note: Low pass filter frequency value is set and returned in kHz
(except for NONE).
:FILTer:FREQuency 15
Note: High pass filter frequency value is set and returned in kHz
(except for NONE).
:FILTer:HPASs:FREQuency 0.3
PN: 10580-00319 Rev. N