14-3 :DISPlay Subsystem
:DISPlay[:WINDow]:TRACe<Tr>:SELect
Description: Selects the given trace, <Tr>, as the active trace. <Tr> is the trace
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: TRAC1
Default Unit: NA
Example: To set trace 2 as the active trace:
Front Panel Access: Measurement, DMR 2 Analyzer, Active Graph
14-8
number in the range 1 to 4 for DMR 2 Analyzer. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1.
:DISPlay:TRACe2:SELect
PN: 10580-00319 Rev. N
DMR 2 Commands
S412E PM