Anritsu LMR Master S412E Programming Manual page 549

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

PTC Commands
[:SENSe]:POWer[:RF]:RANGe[:IMMediate]
Description: Turns off auto ranging and adjusts the receiver reference level once.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To adjust range:
Front Panel Access:Amplitude, Adjust Rx Range
[:SENSe]:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Description: Turns auto range for the receiver on or off. When auto range is on, the
Cmd Parameters: OFF|ON|0|1
Query Parameters: NA
Range: OFF|ON|0|1
Default Value: ON or 1
Default Unit: NA
Example: To turn auto ranging off:
Front Panel Access: Amplitude, Auto Rx Range
S412E PM
In PTC Analyzer measurement, this command adjusts the receiver
reference level of the spectrum graph.
:POWer:RANGe
reference level is automatically adjusted to the proper value to show the
trace on the screen. If the auto ranging is turned off, the reference level
will not adjust according to where the trace is. In PTC Analyzer
measurement, this command adjusts the reference level of the spectrum
graph.
:POWer:RANGe:AUTO OFF
PN: 10580-00319 Rev. N
15-13 [:SENSe] Subsystem
15-45

Advertisement

Table of Contents
loading

Table of Contents