WASH IN PROCESS indicates that the analyzer started a new
sample analysis cycle, the analysis cycle starts with a wash
sequence which is designed to first flush and clean the
analyzer reaction cell and then to capture a fresh sample for
executing the measurement steps of the analysis.

2210AM Analyzer ONLY

The UNIVERSAL 2210AM Ammonia analyzer has a specific
programmed method for executing the wash and sample
capture; the method starts with a PRE-WASH sequence. The
FIRST step in the sequence allows for evacuating the reaction
cell and allows for hold time or stir time (default to 1 sec), this
delay or stir time is user programmable and can be adjusted
to provide the sample conditioner additional time to acquire,
condition and feed a fresh sample to the analyzer from the
process.
To access the PRE-WASH parameters menu:
Press STOP
MENU
PARAMETERS (+) pg 2
Scroll down to STEP 1 using F4 key (see photos of menu
screens)
The following screen capture of the analyzer pre-wash menu
shows an example in which this delay was changed from the
default of 1 sec to 120 seconds to allow time for the sample
conditioner to collect and feed a new sample to the analyzer.
DEFAULT SETUP
You may modify ONLY STEP 1 STIRTIME if delay is needed to
allow the sample conditioner time to prepare and feed fresh
sample to analyzer.
Example below: STEP 1 stir time modified to 120 seconds.
54 |
Addendum
SETUP
(+) pg 2
ANALYSIS
DO NOT MODIFY ANY OTHER
PARAMETER WITHOUT FIRST
CONSULTING FACTORY
CAUTION
2210AM Measurement
ranges
The 2210AM Rev 3.53+ release combined the
4 measurement ranges:
0-2 mg/L

0-10 mg/L D(EFAULT range at power up)

0-50 mg/L

50-500 mg/L

When user selects a new range the associated default cali-
bration parameters are automatically loaded and the method
is adjusted for the new selected range.