Servicing
Capillary Cleaning and/or Replacement
Figure 7–26. Capillary Plate, Keyway
5. Using 7/64 hex drive, remove four #6-32 socket cap head screws.
6. Pull apart the capillary blocks.
Figure 7–27. Capillary and O-Ring Replace
7-23
Thermo Scientific
450iQ Instruction Manual