Poor Accuracy/Feedback; Poor Detection Limits - Thermo Scientific iCAP 6000 Series Maintenance Manual

Icp-oes spectrometer
Hide thumbs Also See for iCAP 6000 Series:
Table of Contents

Advertisement

iCAP Customer Training Manual

6.2 Poor accuracy/feedback

First we should define accuracy as reproducing the standard value once standardised
Proceed by making a standard as in the previous section (10 ppm Cu, Mn, Cd, Zn, Fe)
and using this as the test solution. Remember we are not defining accuracy for this test
as the ability to read a 1 or 100 ppm standard after standardising on the 10 ppm. Most of
the time that problem is operator-related. As far as we are concerned at this point we
have only one standard to test with. If this simple standard will not repeat back for all
elements, check the pump winding first. Replace it if it is used or collapsed.
Note: Pump windings typically last only a couple of days
Check the method to see if a fast pump rate is used for the flush period. If it is, then
make the flush pump rate the same as the analysis rate and try it again. Inaccuracy
can sometimes be traced to the inability of the pump tubing to recover its shape
after being stretched.
Check the flush time for adequate rinse time. A 30-second rinse time is adequate in
most cases but not if a slow pump rate is being used or if a very long piece of tubing
is used (as with the autosampler probe).
7

Poor detection limits

This problem can be also related to
the poor precision problem discussed
earlier and is usually solved by
approaching it as such. However, if
the loss of intensities is especially
pronounced at lower wavelengths it
may be due to a dirty window
mounted in the Purged Optical Path.
UV burn or a dirty mirror is
characterised by a long term decline
(six months or longer) of intensities.
.