Ci Reagent Gas Flow Module - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

CI Reagent Gas Flow Module

Thermo Scientific
The CI reagent gas flow module controls the flow of CI reagent gas into the ion source and is
located at the back left of the ISQ 7000 mass spectrometer. The module uses electronic
pressure control to precisely set the flow of reagent gas. The flow is programmed using
Xcalibur, which allows you to accurately reproduce flows. To install a CI reagent gas flow
module:
1. Shut down your system by following the steps in
page
109.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. If the ISQ 7000 mass spectrometer is too close to the GC, move them apart so you can
access the right side panel.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Figure 435. Removing the Right Side Panel Screws
5. Slide the right side panel toward the back of the instrument and lift it off.
"Shutting Down the ISQ System"
Right Side Panel Screws
ISQ 7000 Hardware Manual
4
Upgrade Equipment
CI Reagent Gas Flow Module
on
381

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents