Chapter 4 Data Storage and Review
Data Storage Settings
Measurement Read Type
The measurement read type determines when the meter sends measurements to outputs as selected in the setup menu
and when properly connected to the meter. These are set individually for each channel. Refer to the individual module
reference guide to set the measurement read type for each channel.
Computer and Printer Export Settings
Refer back to Chapter 3 Instrument Settings of this reference guide to:
•
set the communication protocol
•
turn on printing
•
select the print format
Data Log Settings
Refer Chapter 3 Instrument Settings to turn on Data Log to enable data storage. Orion VERSA STAR Meters have a
2000-datalog capability.
Datalog Review
This option allows for stored data to be:
•
reviewed
•
printed individually, in a group or all
•
deleted
•
analyzed with general statistical information
The datalog will refl ect a screen capture of all the information at the time of datalogging.
To review datalog information
1.
Press log view (9) key.
2.
Press Select (f3) key.
3.
Follow prompts to review information.
Thermo Scientifi c Orion VERSA STAR Advanced Electrochemistry Benchtop Meter
Example Datalog View for DO Measurement
13