Error Messages

Detector
Data System

Instrument Troubleshooting

The control unit of the GC performs a number of test routines to monitor and
report on the status of the major chromatographic and control functions.
Instrument malfunctioning, due to a component failure or to abnormal operating
condition is identified by the red lighting of the Not Ready/Error LED located on
the GC Status Panel.
When lit, this LED indicates that the injector and oven power has been cut off for
safety reasons and an error message indicating the possible cause of error appears
on the display. See

Error Messages

The following table reports the error messages and the explanation of the relevant
correlated problem
Message
TEMP SHUTDOWN
Loss of carrier
HARDWARE SHUTDOWN
Oven T over limits
Instruction Manual
Check the column efficiency periodically.
Ensure that the detector parameters are correctly set.
Check for eventual detector contamination.
Check that the Data System in use is properly connected.
Error
Table 15-6. Error Messages
This message appears in two cases:
The carrier supply may be missing and the actual pressure value is 0, or
the actual pressure cannot reach the setpoint value.
The oven temperature exceeds the temperature limits (350 °C). Error
may be due to a probe malfunctioning.
Messages.
Description
Chapter 15
Maintenance and Troubleshooting
307