View The Dye Calibration History - Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer User Manual

Hide thumbs Also See for Applied Biosystems SeqStudio Genetic Analyzer:
Table of Contents

Advertisement

View the dye
calibration history
SeqStudio
Genetic Analyzer Instrument and Software User Guide
d. Touch a custom dye, then touch Export.
e. Select a location, then touch Export.
f. Import the custom dye on another instrument (
and Service4 Calibration4Dye calibration4Custom
dye4Manage4Import).
In the home screen:
1. Touch
Settings4Maintenance and Service4Calibration history.
Factory calibration (default calibration that is not optimized for the instrument).
1
Manual dye calibration performed by a user.
2
Factory calibration that has been optimized by auto calibration and does not require
3
manual calibration.
Automatic adjustment of the baseline calibration to optimally reduce pull‑up (false
4
secondary peaks under a true peak).
A manual dye calibration and an auto dye calibration have been performed.
5
2. Touch a dye set.
A thumbnail of each capillary calibration spectrum is displayed, with the q, and c
values (see "Spectral Quality Value" on page 196 and "Condition number" on
page 197).
3. Touch
and
to view a full-screen calibration spectrum for each capillary.
4. Touch
to return to the list of dye sets on the Calibration History screen.
5. (Optional) Touch Filter to narrow the dye set list down by dye sets.
a. Select or deselect the dyes listed.
Appendix C Dye calibration and install standard checks
C
Calibrate dyes
Settings4Maintenance
201
1
2
3
1
1
4
5