Measuring
9.3 Evaluation of the test results
The dynamic viscosity (in mPas) is calculated using the
following equation:
= K (
--
) ¡ t
1
2
where:
K = ball constant in mPascm
= density of the ball in g/cm
1
= density of the liquid to be measured at the measuring
2
temperature in g/cm
t = falling time of the ball in seconds.
Test results:
The dynamic viscosity is given in units of mPas (cP) and
must be completed by stating the sample temperature.
The dynamic viscosity may be converted to the kinematic
viscosity by using the following equation:
=
= kinematic viscosity [mm
= dynamic viscosity [mPas]
= density of the liquid sample [g/cm
14
3
/gs (see chapter 11)
3
(see chapter 11)
3
2
2
/s]
[1 mm
/s = 1 cSt]
3
]