Sequence Analysis Results - Thermo Scientific Applied Biosystems SeqStudio Genetic Analyzer User Manual

Hide thumbs Also See for Applied Biosystems SeqStudio Genetic Analyzer:
Table of Contents

Advertisement

Sequence analysis
results
Sequencing results
Result
Quality
SeqStudio
Genetic Analyzer Instrument and Software User Guide
Sequence analysis trace
1
2
3
4
5
Quality Value bars and values:
1
Pure base with QV ³20
Pure base with QV 15–19
Pure base with QV <15
Mixed base
Bases—Mixed bases are highlighted in red. You can set the secondary peak threshold, as a
2
percentage of the primary peak, for consideration as a mixed base by the basecalling
algorithm. Reaching this threshold is a necessary but not sufficient condition for arriving
at a mixed base determination. See "Sequencing settings (base calling)" on page 129.
Trace color hide/show—Touch to open, then touch a color to hide or show.
3
Analyzed trace
4
Thumbnail trace—Drag the center of the pane in the thumbnail trace to display another
5
trace area in the top pane. Drag the right or left handle of the pane to zoom horizontally.
Zoom tools—Touch to open.
6
Next trace tool—Touch to view the raw trace or EPT for the well.
7
QC flags
—All QC tests passed.
—At least 1 warning quality alert was triggered.
—At least 1 failing quality alert was triggered.
Chapter 8 View and analyze results
View results on the instrument
Description
8
6
7
115