OdaLog® L2 – Gas Logger User Manual

6.2. Switching On

Press and hold the button for three seconds ("On" will count down from 3), then release. The LCD will
show (in order):
o Test Screen – All segments visible/black.
o "CLOCK...not Set" (See Note 3).
o Installed OdaLog® Firmware Version.
o Days Since Last Calibrated.
o "STATUS" which displays in order.
o "PEAK" gas level since switch on or log start.
o "CLOCK" (shown in 24-hour time. If the clock is not set (See Section 6.3), this value may be
incorrect on screen and in the instrument log – This must be set in the OdaStat-G™ Software –
refer to the OdaStat-G™ Quick User Guide or "Help" section in the software manual).
o Temperature
o "BATT" (Battery), battery volts
o Logging "MEMORY"- % of memory used.
o "TYPE" of logging - "roLL" (continue logging after the memory is full), "oncE" (stop logging after
the memory is full).
o Logging interval, the time between gas readings displayed as "SECONDS" (set in the OdaStat-G™
software).
o "CALGAS", Shows Calibration Gas Level in ppm as set via the OdaStat-G™ software package.
o "CLOCK not SEt" (see note 3 & Section 6.3).
o "LOG FULL" (see note 4 & section 6.7).
The display will then revert to display instantaneous gas readings in ppm.
Note 1: Releasing the button during the countdown routine will cause an exit to the "OFF" screen.
Note 2: "Peak" is constantly updated when OdaLog® is in use. The Peak value is reset at the switch on
and at the start of each logging session.
Note 3: If the clock is not set, "CLOCK not SEt" will be displayed. The message will remain on the display
until the button is pressed to continue (see Section 6.3).
Note 4: If the Logger memory is full and you have "Auto Log Start" & "Stop log when full" active (refer to
the "Instrument Parameters" section in the OdaStat-G™ software package), the unit will not start
logging, and the display will show "LOG", "FULL", and the LED" s will illuminate at the same time.
The message will remain on the display until the button is pressed to continue.

6.3. "CLOCK not SEt" Message

If the clock has not been set since the last battery change, "CLOCK not...SEt" will be displayed upon start-
up and when starting logging functions. This message will be displayed until the button is pressed. Logging
will start normally; however, the time and date will be incorrect.
P/N: 10-1000

Figure 6: Clock not SEt

Revision G, January 2024