Chapter 5 Create or modify a plate setup from the Plate Manager
Factory-installed items cannot be edited or deleted. To create a new item from a
factory-installed item, copy, edit, then save the new item.
1. In the Properties tab, click Setting Details, or select Actions4Manage analysis
settings.
2. To create new analysis settings:
a. Select the default analysis settings or user-created analysis settings, then
click Copy.
b. Enter a name and edit settings as needed (see "Fragment/HID analysis
settings (size calling)" on page 126 or "Sequencing settings (base calling)" on
page 129).
c. Click Save.
3. As needed, select an analysis setting of interest, then click Edit or Delete
(user-created settings only).
4. Click Close.
The file name convention determines how the data files (AB1 or FSA) associated with
a plate are named.
The default file name convention is:
<well>_<sample name>_<sample type>_<date and timestamp>.fsa
<well>_<sample name>_<date and timestamp>.ab1
<well>_<sample name>_<amplicon>_<specimen>_<date and
timestamp>.ab1
To change the default settings, in the Properties tab:
1. Select Actions4Edit the file name convention.
2. Click File attributes, then select the attributes to include in the data file name.
Note: The timestamp attribute cannot be deselected; it is always included in the
data file name.
Additional plate settings (Plate Manager)
Default settings
5