Calibration
Calibration At Custom
Concentrations
Using DKA
For Maximum Accuracy
Thermo Scientific Orion 2117HL High Level Chloride Analyzer User Guide
The Thermo Scientific Orion 2117HL chloride analyzer software has
default values to accommodate a 0.50 mL addition of standard 1,
three 0.50 mL additions of standard 2 and flow cell volume of
approximately 95 mL.
The operator has the option to use alternate standards by changing, in
the setup mode, the flow cell volume, the standard 1 and standard 2
concentration values and the standard 1 and standard 2 volume values to
be used.
Let V
= Flow cell volume
C
C
= Standard 1 concentration
1
C
= Standard 2 concentration
2
V
= Volume of standard 1 added
1
V
= Volume of standard 2 added
2
dC
= Concentration change due to the first addition
1
dC
= Concentration change due to the second addition
2
dC
= C
x V
/ (V
1
1
1
dC
= C
x V
/ (V
2
2
2
dC
should be at least twice the normal concentration in the sample
1
dC
should be about ten times of dC
2
dC
should not be less than 10 ppb
1
Default: 95.0 mL
Default: 2.96 %
Default: 2.96 %
Default: 0.50 mL
Default: 1.50 mL
+ V
)
1
C
+ V
+ V
)
1
2
C
1
IV-13