From Data base wizard select Restore and
carry out the same procedure as Backup
database above

5 Maintenance

5.1 Introduction

The iCAP 6000 has been designed for minimum maintenance; however the sample introduction
components should be checked regularly for contamination and wear.

5.2 User Maintenance

Routine operator maintenance of the iCAP 6000 is mainly concerned with keeping the instrument
clean.
5.2.1 Instrument Cleaning
'Before using any cleaning or decontamination methods except those specified by the
manufacturer, USERS should check with the manufacturer that the proposed method will not
damage the equipment.'
Warning: The iCAP 6000 instrument covers are made of ABS plastic materials,
which can be damaged by strong solvents and concentrated acids.
Any spillage on the external covers or within the sample introduction areas should be cleaned up
immediately, using appropriate safety precautions.
Stains and marks on the covers should be removed with a soft cloth moistened with a dilute
detergent solution. Do not use any solvent based cleaners.
5.2.2 Sample Introduction System Cleaning and Decontamination
Failure to maintain the sample introduction system can result in erroneous results, poor precision,
poor detection limits and blockages.
After use, the instrument shut down procedures should be followed.
Components contaminated with sample residues should be cleaned.
It is recommended that several spares for each part are available as blockages, sample
contamination and breakages often happen at critical moments during analysis.
Suitable protective clothing, glasses and gloves should be worn.
5.2.2.1
Torch Cleaning
Reverse the torch assembly process detailed previously.
Warning: Allow at least 10 minutes for any hot components to cool before removing
them from the torch compartment. Care should be taken to remove any broken
glass from the Duo radial POP window, if a breakage occurs in the torch box.