Return (In addition to the results of "Input Tests Common to all XPS
Functions")
• 0:
No error.
• -17:
Parameter out of range or incorrect.
• -22:
Not allowed action.
• -25:
Following Error.
• -26:
Kill command or Emergency signal: check each positioners and each
slave positioners, check that motion does not exceed software limits
when combined with mapping and other features.
• -27:
Move Aborted.
• -28:
Home search timeout.
• -33:
Motion done timeout.
• -35:
Position is outside of travel limits.
• -44:
Slave error disabling master.
• -49:
Inconsistent mechanical zero during home search.
• -50:
Motor initialization error. Check InitializationAccelerationLevel,
ScalingAcceleration, MaximumJerkTime, EncoderResolution or
EncoderScalePitch.
• -113:
Both ends of run activated.
• -120:
Warning following error during move with position compare enabled.
• -200:
Invalid socket.
• -205:
Not enable in your configuration.