ports A and T1). If the calibration is not completed correctly, these lines
abort the program.
Lines 1130 to 1210
Execute the calibration of channels 2 to 4 in the same way.
Line 1260
Displays the message that prompts for connecting a DUT (Device Under
Test) and waits for the OK button to be clicked after the connection.
Lines 1280 to 1290
Generate a trigger to start a single sweep and wait until the measurement
finishes (1 is read out with the SCPI.IEEE4882.OPC object).
Lines 1310 to 1330
Execute auto scale for trace 1 of channels 1 to 4.
Line 1350
Displays the message asking whether you want to perform measurement
again.
Line 1360
If the Yes button is clicked, returns to the DUT connection section.
Function procedure: ECal_solt (lines 1410 to 1770)
Line 1460
Clears the error queue.
Lines 1460 to 1480
Display the message that prompts for connecting the Tset_Port of the
E5091A to the ECal module and wait for the OK button to be clicked after
the connection.
Line 1500
Enables the error handling routine starting from Ecal_Err (lines 1670 to
1740). If a runtime error occurs, the program goes to the error handling
routine.
Line 1540
If solt is 1, executes the ECal command that performs 1-port calibration on
port Ana_port(0) of channel Ch.
Line 1560
If solt is 2, executes the ECal command that performs full 2-port
calibration on port Ana_port of channel Ch.
Line 1580