Executing Power Calibration
Overview
Program Overview
Program Description
Other topics about Application Programs
Overview
The pow_cal.vba shows a sample program (VBA program) for executing
power calibration using the E4418B power meter and the E4412A power
sensor. This VBA program consists of the following modules:
Object
name
mdlPowCal
Module1
Module2
Program Overview
The program connects the E5071C and the E4418B (GPIB address: 13)
through the USB/GPIB interface as shown below. Then, the program
executes the zero adjustment and calibration of the power sensor
(E4412A) connected to the power meter (E4418B) as necessary. Finally, it
executes the power calibration of the E5071C and saves the obtained
power calibration data array into a file.
Connection between E5071C and power meter
Program Description
Module type
Standard
module
Standard
module
Description
Executes power calibration
Definition files for using VISA
library
Programming
1429