ESR
Up to 15 instruments can be connected
The total cable length is restricted to a maximum of 15 m or 2 m times the number
of devices, whichever is less; the cable lenth between two instruments should not
exceed 2 m.
A wired "OR"-connection is used if several instruments are connected in parallel.
Any connected IEC-bus cables should be terminated by an instrument or controller.
GPIB Interface Messages
Interface messages are transmitted to the instrument on the data lines, with the atten-
tion line (ATN) being active (LOW). They are used for communication between the con-
troller and the instrument and can only be sent by a computer which has the function of
a GPIB bus controller. GPIB interface messages can be further subdivided into:
Universal commands: act on all instruments connected to the GPIB bus without
previous addressing
Addressed commands: only act on instruments previously addressed as listeners
The following figure provides an overview of the available communication lines used by
the GPIB interface.