ESR
Range Log 5 dB ← Range
Sets the level display range to 5 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see
SPACing
Display range:
DISP:WIND:TRAC:Y 5DB, see
on page 844
Range Log 1 dB ← Range
Sets the level display range to 1 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see
SPACing
Display range:
DISP:WIND:TRAC:Y 1DB, see
on page 844
Range Log Manual ← Range
Opens an edit dialog box to define the display range of a logarithmic level axis man-
ually.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see
SPACing
Display range:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
Range Linear % ← Range
Selects linear scaling for the level axis in %.
The grid is divided into decadal sections.
Markers are displayed in the selected unit ("Unit" softkey). Delta markers are displayed
in % referenced to the voltage value at the position of marker 1. This is the default set-
ting for linear scaling.
Remote command:
DISP:TRAC:Y:SPAC LIN, see
on page 668
Range Lin. Unit ← Range
Selects linear scaling in dB for the level display range, i.e. the horizontal lines are
labeled in dB.
Markers are displayed in the selected unit ("Unit" softkey). Delta markers are displayed
in dB referenced to the power value at the position of marker 1.
User Manual 1175.7068.02 ─ 12
Softkeys of the Amplitude Menu in I/Q Analyzer Mode
on page 668
on page 668
on page 668
DISPlay[:WINDow<n>]:TRACe<t>:Y:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
DISPlay[:WINDow<n>]:TRACe<t>:Y:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
DISPlay[:WINDow<n>]:TRACe<t>:Y:
on page 844
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
I/Q Analyzer