R&S ESR Series User Manual page 435

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Figure 6-22: Fundamental wave and the frequency and level reference point
5. Make the step size for the center frequency equal to the signal frequency
a) Press the [FREQ] key.
b) Press the "CF-Stepsize" softkey and press the "= Marker" softkey in the sub-
The step size for the center frequency is now equal to the marker frequency.
6. Set the center frequency to the second harmonic of the signal.
a) Press the [FREQ] key.
b) Press the [UPARROW] key once.
7. Place the delta marker on the second harmonic.
a) Press the [MKR ->] key.
b) Press the "Peak" softkey.
The delta marker moves to the maximum of the second harmonic. The displayed
level result is relative to the reference point level (= fundamental wave level).
User Manual 1175.7068.02 ─ 12
menu.
The center frequency is set to the second harmonic.
Spectrum Measurements
Advanced Measurement Examples
432