R&S ESR Series User Manual page 383

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Remote command:
[SENSe:]SWEep:EGATe:HOLDoff
Gate Length (Gate Mode Edge) ← Gate Settings
Opens an edit dialog box to enter the gate length. The gate length in relation to the
sweep is indicated by a line labeled "GL".
The length of the gate signal defines if the sweep is to be interrupted. Only in the edge-
triggered mode the gate length can be set, while in the level-triggered the gate length
depends on the length of the gate signal.
For details also see
Remote command:
[SENSe:]SWEep:EGATe:LENGth
Trg/Gate Source ← Gate Settings
See
Trg/Gate Level ← Gate Settings
See
Trg/Gate Polarity ← Gate Settings
See
Sweep Time ← Gate Settings
Opens an edit dialog box to change the sweep time in order to obtain a higher resolu-
tion for positioning gate delay and gate length. When leaving the "Gate Settings" sub-
menu, the original sweep time is retrieved.
For details also see
Res BW Manual ← Gate Settings
Opens an edit dialog box to enter a value for the resolution bandwidth. The available
resolution bandwidths are specified in the data sheet.
For details on the correlation between resolution bandwidth and filter type refer to
Chapter 6.2.6.3, "Selecting the Appropriate Filter
Numeric input is always rounded to the nearest possible bandwidth. For rotary knob or
UP/DNARROW key inputs, the bandwidth is adjusted in steps either upwards or down-
wards.
The manual input mode of the resolution bandwidth is indicated by a green bullet next
to the "RBW" display in the channel bar.
When measuring Spurious Emissions, using this softkey automatically opens the
"Sweep List" dialog (see
Remote command:
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
[SENSe:]BANDwidth|BWIDth[:RESolution]
User Manual 1175.7068.02 ─ 12
Chapter 6.2.8.3, "Using Gated Sweep
"Trg/Gate Source"
on page 375.
"Trg/Gate Level"
on page 377.
"Trg/Gate Polarity"
on page 377.
Chapter 6.2.8.3, "Using Gated Sweep
"Sweep List dialog box"
on page 841
on page 842
Type",
on page 366.
on page 289).
on page 830
on page 670
Spectrum Measurements
Configuration
Operation",
on page 381.
Operation",
on page 381.
380