<generator>
Parameters:
<Value>
Example:
Mode:
Manual operation:
SOURce<n>:EXTernal<generator>:FREQuency:OFFSet<m> <Offset>
This command defines the frequency offset of the selected generator with reference to
the receive frequency.
This command is only valid for the External Tracking Generator option (R&S FSV-B10).
Suffix:
<n>
<generator>
<m>
Parameters:
<Offset>
Example:
Mode:
User Manual 1175.7068.02 ─ 12
Remote Commands to Control the Tracking Generator
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
<numeric value>
*RST:
1
SOUR:EXT:FREQ:NUM 4"
"SOUR:EXT:FREQ:DEN 3"
Sets a multiplication factor of 4/3, i.e. the transmit frequency of
the generator is 4/3 times the analyzer frequency.
A, NF
See
"Generator Frequency (Numerator/Denominator/Offset)"
on page 489
.
irrelevant
1 | 2
tracking generator
For Noise Figure measurements (K30): 1 only
1 | 2
offset (for Noise Figure measurements (K30) only)
<numeric value>, specified in Hz, kHz, MHz or GHz, rounded to
the nearest Hz
*RST:
0 Hz
Select the multiplication factor such that the frequency range of
the generator is not exceeded if the following formula is applied
to the start and stop frequency of the analyzer:
For Noise Figure measurements (R&S FSV-K30), the following
formula applies:
Generator Frequency = [(LO + Offset 1) * Factor 1 / Factor 2)] +
Offset 2
SOUR:EXT:FREQ:OFFS 1GHZ
Sets a frequency offset of the generator transmit frequency com-
pared to the analyzer receive frequency of 1 GHz.
A, NF