[SENSe:]LIST:RANGe<range>:LIMit:STARt <Limit>
This command sets an absolute limit value at the start frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Suffix:
<range>
Parameters:
<Limit>
Example:
Manual operation:
[SENSe:]LIST:RANGe<range>:LIMit:STATe
This command activates or deactivates the limit check for all ranges.
Suffix:
<range>
Parameters:
<State>
Example:
Manual operation:
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
.
1...20
range
ON | OFF
*RST:
OFF
LIST:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
See
"Preamp"
on page 290
.
1...20
range
-400 to in 400 dBm
*RST:
13 dBm
LIST:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
See
"Abs Limit Start"
on page 291
.
irrelevant
ON | OFF
*RST:
OFF
LIST:RANG3:LIM:STAT ON
Activates a limit check for all ranges.
See
"Limit Check 1-4"
on page 291
Remote Control
755