'y meas. unit
'--------- Trace and detector setting --------------------------------------
CALL InstrWrite(analyzer,"DISP:TRAC1:MODE AVER")
'Trace1 average
CALL InstrWrite(analyzer,"AVER:TYPE VID")
'Average mode video; "LIN" for linear
CALL InstrWrite(analyzer,"SWE:COUN 10")
'Sweep count
CALL InstrWrite(analyzer,"DISP:TRAC2 OFF")
'Trace2 blank
CALL InstrWrite(analyzer,"DISP:TRAC3 OFF")
'Trace3 blank
CALL InstrWrite(analyzer,"DISP:TRAC4 OFF")
'Trace4 blank
CALL InstrWrite(analyzer,"DISP:TRAC5 OFF")
'Trace5 blank
CALL InstrWrite(analyzer,"DISP:TRAC6 OFF")
'Trace6 blank
CALL InstrWrite(analyzer,"CALC:MATH:STAT OFF")
'Trace difference off
CALL InstrWrite(analyzer,"DET1 RMS")
'Detector Trace1
CALL InstrWrite(analyzer,"DET2:AUTO ON")
'Detector Trace2
CALL InstrWrite(analyzer,"DET3:AUTO ON")
'Detector Trace3
CALL InstrWrite(analyzer,"DET4:AUTO ON")
'Detector Trace4
CALL InstrWrite(analyzer,"DET5:AUTO ON")
'Detector Trace5
CALL InstrWrite(analyzer,"DET6:AUTO ON")
'Detector Trace6
'--------- Bandwidths and sweep time ---------------------------------------
CALL InstrWrite(analyzer,"BAND:RES 100KHz")
'Resolution bandwidth (*)
CALL InstrWrite(analyzer,"BAND:VID 1MHz")
'Video bandwidth
CALL InstrWrite(analyzer,"SWE:TIME 100ms")
'Sweep time
END SUB
Using Markers and Delta Markers
The markers are used for marking points on traces, reading out measurement results
and for selecting a display area quickly.
Marker Search Functions, Restricting the Search Range
The following example is based on an AM-modulated signal at 100 MHz that has the
following characteristics:
User Manual 1175.7068.02 ─ 12
(*)
(*)
(*)
(*)
(*)
(*)
(*)
(*)
Remote Control
Controlling the R&S ESR Remotely