R&S ESR Series User Manual page 346

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
6.2.1.2
Initial Configuration
The initial configuration is selected such that the RF input is always protected against
overload, provided that the applied signal levels are in the allowed range for the instru-
ment.
The parameter set of the initial configuration can be customized using the "Startup
Recall" softkey in the "Save/Rcl" menu. For further information refer to
"Saving and Recalling Settings Files – SAVE/RCL
Table 6-5: Initial configuration
Parameter
mode
sweep mode
center frequency
center frequency step size
span
RF attenuation
reference level
level range
level unit
sweep time
resolution bandwidth
video bandwidth
FFT filters
span/RBW
RBW/VBW
sweep
trigger
trace 1
trace 2/3/4/5/6
detector
frequency offset
reference level offset
reference level position
grid
cal correction
User Manual 1175.7068.02 ─ 12
Spectrum Measurements
Key",
on page 529.
Setting
Spectrum
auto
f
/2
max
0.1 * span
maximum possible (depends on instrument model)
0 dB
-10 dBm
100 dB log
dBm
auto
auto (3 MHz)
auto (3 MHz)
off
100
1
cont
free run
clr write
blank
auto peak
0 Hz
0 dB
100 %
abs
on
Configuration
Chapter 10.1,
343