The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
<m>
Parameters:
<State>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
This command queries the results for the third order intercept point measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
<m>
Return values:
<TOI>
Example:
Usage:
Manual operation:
11.5.1.11
Measuring Statistics
The following commands configure and perform statistics measurements.
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
.
Selects the measurement window.
irrelevant
ON | OFF
*RST:
OFF
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
See
"TOI"
on page 224
See
"TOI"
on page 335
.
Selects the measurement window.
irrelevant
Third order intercept point.
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES?
Outputs the measured value.
Query only
See
"TOI"
on page 224
See
"TOI"
on page 335
Remote Control
791