ESR
Time ← Trg/Gate Source
Opens an edit dialog box to define a repetition interval in which the measurement is
triggered. The shortest interval is 2 ms.
Remote command:
TRIG:SOUR
Trigger Level
Defines the trigger level as a numeric value.
In the trigger mode "Time", this softkey is not available.
Remote command:
TRIGger<n>[:SEQuence]:LEVel:IFPower
TRIGger<n>[:SEQuence]:LEVel:VIDeo
Trigger Polarity
Sets the polarity of the trigger source.
The sweep starts after a positive or negative edge of the trigger signal. The default set-
ting is "Pos". The setting applies to all modes with the exception of the "Free Run" and
"Time" mode.
"Pos"
"Neg"
Remote command:
TRIGger<n>[:SEQuence]:SLOPe
[SENSe:]SWEep:EGATe:POLarity
Trigger Offset
Opens an edit dialog box to enter the time offset between the trigger signal and the
start of the sweep.
offset > 0:
offset < 0:
In the "External" or "IF Power" trigger mode, a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger delay (trig-
ger offset) as well.
Tip: To determine the trigger point in the sample (for "External" or "IF Power" trigger
mode), use the
In the "Time" trigger mode, this softkey is not available.
Remote command:
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]
User Manual 1175.7068.02 ─ 12
Softkeys of the Trigger Menu in I/Q Analyzer Mode
TIMETRIGger<n>[:SEQuence]:SOURce
Level triggering: the sweep is stopped by the logic "0" signal and
restarted by the logical "1" signal after the gate delay time has
elapsed.
Edge triggering: the sweep is continued on a "0" to "1" transition for
the gate length duration after the gate delay time has elapsed.
Start of the sweep is delayed
Sweep starts earlier (pre-trigger)
Only possible for span = 0 (e.g. I/Q Analyzer mode) and gated trigger
switched off
Maximum allowed range limited by the sweep time:
pretrigger
max
TRACe<n>:IQ:TPISample?
on page 839
on page 838
on page 838
on page 839
on page 842
= sweep time
command.
on page 836