R&S ESR Series User Manual page 851

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Example:
Manual operation:
CALCulate<n>:MATH:MODE <Method>
This command selects the method for the trace math calculations.
For details see
Suffix:
<n>
Parameters:
<Method>
Example:
Manual operation:
CALCulate<n>:MATH:POSition <Position>
This command defines the position of the result of the trace mathematics. The indica-
tion is in % of the screen height, with 100 % corresponding to the upper diagram bor-
der.
Suffix:
<n>
Parameters:
<Position>
Example:
Manual operation:
CALCulate<n>:MATH:STATe <State>
This command switches the mathematical relation of traces on or off.
Suffix:
<n>
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
(TRACe1-TRACe5)
Subtracts trace 5 from trace 1.
(TRACe1-TRACe6)
Subtracts trace 6 from trace 1.
CALC1:MATH (TRACe1 – TRACe2)
Selects the subtraction of trace 2 from trace 1.
See
"Trace Math"
"Trace Math Mode"
on page 394.
.
irrelevant
LINear | LOGarithmic | POWer
*RST:
LOG
CALC:MATH:MODE LIN
Selects linear averaging for trace math calculations.
See
"Lin"
on page 394
See
"Log"
on page 395
See
"Power"
on page 395
.
Selects the measurement window.
-100PCT to 200PCT
*RST:
50PCT
CALC:MATH:POS 50PCT
Sets the position to the horizontal diagram center.
See
"Trace Math Position"
.
irrelevant
on page 394
on page 395
Remote Control