Example:
Manual operation:
[SENSe:]FMEasurement:LISN[:TYPE] <Type>
This command turns automatic control of the network used for the final measurement
on and off. It also selects the type of network in use.
Parameters:
<Type>
Example:
Manual operation:
INPut:LISN:FILTer:HPAS[:STATe] <State>
This command turns the 150 kHz highpass filter of the ENV216 network for the pre-
measurement on and off.
User Manual 1175.7068.02 ─ 12
L2
Available for networks with four phases (R&S ESH2Z5,
R&S ENV4200 and R&S ENV432)
L3
Available for networks with four phases (R&S ESH2Z5,
R&S ENV4200 and R&S ENV432)
N
*RST:
L1
FME:LISN:PHAS L1,N
See
"Final Test Phase"
ENV216
R&S ENV 216: two phases and highpass are controllable.
ENV432
R&S ENV 432: four phases are controllable.
ENV4200
R&S ENV 4200: four phases are controllable.
ESH2Z5
R&S ESH2-Z5: four phases and protective earth are controlla-
ble.
ESH3Z5
R&S ESH3-Z5: two phases and protective earth are controllable.
FOURphase
R&S ESH2-Z5: four phases and protective earth are controlla-
ble.
OFF
Turns off remote control of the LISN.
TWOPhase
R&S ESH3-Z5: two phases and protective earth are controllable.
*RST:
OFF
FME:LISN TWOP
Turns automatic control of the network on and selects the R&S
ESH3-Z5 network.
See
"LISN Type"
on page 197
Remote Commands in Receiver Mode
on page 197
Remote Control